CAPSURE SP
Report
- Report Number
- 2649622-2011-13918
- Event Type
- Death
- Date Received
- September 26, 2011
- Date of Event
- June 8, 1997
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S15
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SINGLE CHAMBER PACEMAKER AND LEAD ASSOCIATED WITH THIS ADVERSE OUTCOME WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) -THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED TO THE MANUFACTURER AND ANALYZED. NO ANOMALIES WERE FOUND. THE OUTER INSULATION SHOWED COSMETIC DEPRESSION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SINGLE CHAMBER PACEMAKER AND LEAD ASSOCIATED WITH THIS ADVERSE OUTCOME WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. EVALUATION SUMMARY: (B)(4) -THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED TO THE MANUFACTURER AND ANALYZED. NO ANOMALIES WERE FOUND. THE OUTER INSULATION SHOWED COSMETIC DEPRESSION. (B)(4) - THE DEVICE WAS RETURNED TO THE MANUFACTURER AND ANALYZED. THE DEVICE EXPERIENCED NORMAL DEPLETION, AND MET EXPECTED LONGEVITY.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SINGLE CHAMBER PACEMAKER AND LEAD ASSOCIATED WITH THIS ADVERSE OUTCOME WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
A SINGLE CHAMBER IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS RETURNED TO THE MANUFACTURER FROM AN UNKNOWN SOURCE WITH NO INFORMATION. FURTHER REVIEW OF THE MANUFACTURER'S DATABASE INDICATED THE PATIENT DIED APPROXIMATELY ONE MONTH POST THE DEVICE SYSTEM IMPLANT. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5024M | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Death |