FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 22644385 · Received July 29, 2025

Report

Report Number
1911916-2025-00545
Event Type
Malfunction
Date Received
July 29, 2025
Date of Event
July 23, 2025
Report Date
August 8, 2025
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
00382903096534
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP FOR DEVICE EVALUATION. A REPORT WAS RECEIVED INDICATING THE PRESENCE OF AN EXTRANEOUS PIECE OF PLASTIC ON THE LUER LOCK TIP OF A SYRINGE. TO SUPPORT THE INVESTIGATION, ONE SAMPLE¿INCLUDING THE PACKAGING BLISTER TOP WEB¿WAS SUBMITTED FOR EVALUATION BY THE QUALITY TEAM. UPON VISUAL INSPECTION, A MOLDING FLASH APPROXIMATELY 1/16" IN LENGTH WAS OBSERVED ON THE SYRINGE BARREL LUER TIP. NO ADDITIONAL DEFECTS OR IRREGULARITIES WERE IDENTIFIED. THIS CONDITION CAN OCCUR DURING THE MOLDING PROCESS IF THERE IS A DELAY IN PART EJECTION. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR MATERIAL NUMBER 309653, LOT 5142409. THE REVIEW FOUND NO QUALITY ISSUES DURING PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. NO RELATED QUALITY NOTIFICATIONS WERE IDENTIFIED. ALL MANUFACTURING PROCESSES AND FINAL INSPECTIONS WERE PERFORMED IN ACCORDANCE WITH SPECIFICATION REQUIREMENTS. THE MOLDING PROCESS PARAMETERS WERE VERIFIED AND FOUND TO BE WITHIN SPECIFICATION, WITH NO FLASH OBSERVED ON PARTS DURING PRODUCTION. THE RETURNED SAMPLE WILL BE SHARED WITH PRODUCTION ASSOCIATES TO RAISE AWARENESS. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND ANALYSIS OF THE RETURNED SAMPLE, THE CUSTOMER-REPORTED CONDITION HAS BEEN CONFIRMED.

Additional Manufacturer Narrative · 0

H11: A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT IS REPORTED FOREIGN MATTER. EVENT DETAILS: MATERIAL#: 309653, BATCH#: 5142409. ¿ON (B)(6) 2025, WHILE MIXING INSIDE THE CLEAN ROOM, WE FOUND 1 BD SYRINGE 50CC REF#: 309653, LOT#: 5142409, EXPIRY DATE: 2030/04/30 WITH AN EXTRA PIECE OF PLASTIC ON THE LUER LOCK TIP OF THE SYRINGE.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
887567 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BECTON DICKINSON 5142409 00382903096534

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown