FDA Adverse Event Malfunction Summary report: N

WARMTOUCH COMPACT WARMING UNIT

MDR report key: 2264177 · Received September 2, 2011

Report

Report Number
2936999-2011-00556
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 5, 2011
Report Date
August 5, 2011
Manufacturer
COVIDIEN, FORMERLY TYCOHEALTHCARE
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE WT-5900 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5900 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND IS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5900 WARMTOUCH IS K020604. THE REPORTED CUSTOMER COMPLAINT IS CONFIRMED. THE TEMPERATURE CONTROL WAS FOUND TO BE DEFECTIVE AND THE FAULT ISOLATED TO A DEFECTIVE CIRCUIT BOARD. ADDITIONALLY, THE FILTER COVER WAS REPLACED. REPLACED FILTER COVER AND DEFECTIVE CIRCUIT BOARD. PARTS REPLACED AND UNIT PASSED ALL FUNCTIONAL TESTING. INFORMATION HAS BEEN ADDED TO THE DATABASE FOR TRENDING PURPOSES.

Description of Event or Problem · 1

THE COMPANY RECEIVED A REPORT WHERE IT WAS CLAIMED THAT THE FILTER COVER MELTED DUE TO THE UNIT BEING HOT. THE CALLER COULD NOT RECALL THE TEMPERATURE OF THE DEVICE. UNIT IN USE WITH A PATIENT, THERE WAS NO PATIENT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WARMTOUCH COMPACT WARMING UNIT CONVECTIVE AIR WARMER DWJ COVIDIEN, FORMERLY TYCOHEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1