MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2025-08243
- Event Type
- Injury
- Date Received
- July 28, 2025
- Date of Event
- June 4, 2025
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000549
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON SEPTEMBER 16, 2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON SEPTEMBER 23, 2025, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR THEN CONDUCTED VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE DETERMINED THAT THE BREAST IMPLANT WAS FOUND TO BE RUPTURED AND RECEIVED IN THREE PARTS. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN AN AREA OF THE TEAR ON THE ANTERIOR VIEW, MEASURING APPROXIMATELY 0.2 CM. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. THE CAUSE OF THE RUPTURE IN THE REMAINING AREA OF THE TEAR COULD NOT BE IDENTIFIED. MENTOR ALSO PERFORMED A MANUFACTURING RECORD EVALUATION RELATED TO THE REPORTED COMPLAINT FOR THE FINISHED DEVICE LOT NUMBER. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED AS PART OF THIS EVALUATION. REGARDING THE REPORTED CAPSULAR CONTRACTURE CONDITION, THERE ARE NOT ENOUGH DETAILS TO DETERMINE THE FACTORS THAT MAY HAVE CAUSED THIS CONDITION. CAPSULAR CONTRACTURE IN THE PATIENT¿S BREAST MIGHT BE THE RESULT OF THE BODY¿S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. FURTHER, CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE CONTRALATERAL DEVICE WAS ALSO RECEIVED (LOT NUMBER 9760364). THE PATIENT DID NOT REPORT ANY ISSUE WITH THIS DEVICE. THEREFORE, NO FURTHER ANALYSIS OF THE CONTRALATERAL DEVICE IS REQUIRED. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE, CAPSULAR CONTRACTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH A 550CC MENTOR MEMORYGEL BREAST IMPLANT ON THE RIGHT BREAST. POST-OPERATIVELY, THE PATIENT SUFFERED RIGHT BREAST IMPLANT RUPTURE AND RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE UNSPECIFIED). AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL ON (B)(6) 2025. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1412343 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 9753913 | 00081317000549 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR | Female | Required Intervention |