STEALTHSTATION¿ S8 SYSTEM
Report
- Report Number
- 1723170-2025-02804
- Event Type
- Malfunction
- Date Received
- July 28, 2025
- Date of Event
- July 1, 2025
- Report Date
- December 19, 2025
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- UDI-DI
- 00763000272739
- PMA / PMN Number
- K162309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
H3, H6; THE CAMERA, LOT NUMBER: P903578, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE RETURNED PSU HAD SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY DATING BACK TO (B)(6) 2020, AND INTERMITTENT ILLUMINATOR CURRENT LOW DATING BACK TO (B)(6) 2020. THERE WAS A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT 1.102MM WITH A PASSING THRESHOLD OF .250MM. THIS PSU HAD NOT BEEN PREVIOUSLY RETURNED FOR A FAILURE. CODES B01, C02, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
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
CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821R, H6: NO PRODUCTS HAVE BEEN RETURNED TO MEDTRONIC FOR ANALYSIS. CODES B17, C20, AND D15 ARE APPLICABLE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED INTRA-OPERATIVELY DURING A CRANIAL RESECTION PROCEDURE. IT WAS REPORTED THAT THE SYSTEM WAS SHOWING "LOCALIZER FAULTED". THE SYSTEM WAS REBOOTED 5X WITHOUT ANY CHANGE. THERE WAS A LESS THAN ONE HOUR SURGICAL DELAY AND NO IMPACT ON PATIENT OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 610722 | STEALTHSTATION¿ S8 SYSTEM | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC. | 9735665 | 00763000272739 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Male | "SEE H11...." |