FDA Adverse Event Malfunction Summary report: N

FARADRIVE STEERABLE SHEATH CLEAR

MDR report key: 22631454 · Received July 28, 2025

Report

Report Number
2124215-2025-50475
Event Type
Malfunction
Date Received
July 28, 2025
Date of Event
June 30, 2025
Report Date
November 6, 2025
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
DRA
UDI-DI
00191506043131
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE TECHNICAL ANALYSIS: VISUAL INSPECTION REVEALED NO OUTER ABNORMALITIES WERE NOTED. THE SHEATH WAS LEAK TESTED, AND THE SHEATH PASSED ALL LEAK TESTING. THE VALVES WERE INSPECTED USING MICROSCOPY. NO SIGNIFICANT DAMAGE WAS NOTED, AND NO ABNORMALITIES WERE OBSERVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT FARADRIVE STEERABLE SHEATH CLEAR WAS SELECTED FOR PULMONARY VEIN ISOLATION. DURING PREPARATION NO ABNORMALITIES WAS NOTED DURING PREPARATION OF THE SHEATH. IT WAS MENTIONED THAT AIR CONTINUED TO ASPIRATED FROM THE SHEATH. THUS THE SHEATH WAS REPLACED WITH SAME MODEL SHEATH AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. A SYRINGE PUMP OF NON BOSTON SCIENTIFIC PUMP IRRIGATION METHOD WAS USED WITH A FLOW RATE OF 30ML/H. NO PATIENT COMPLICATION WAS REPORTED. THE DEVICE IS EXPECTED TO BE RETURN FOR ANALYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT FARADRIVE STEERABLE SHEATH CLEAR WAS SELECTED FOR PULMONARY VEIN ISOLATION. DURING PREPARATION NO ABNORMALITIES WAS NOTED DURING PREPARATION OF THE SHEATH. IT WAS MENTIONED THAT AIR CONTINUED TO ASPIRATED FROM THE SHEATH. THUS THE SHEATH WAS REPLACED WITH SAME MODEL SHEATH AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. A SYRINGE PUMP OF NON BOSTON SCIENTIFIC PUMP IRRIGATION METHOD WAS USED WITH A FLOW RATE OF 30ML/H. NO PATIENT COMPLICATION WAS REPORTED. THE DEVICE IS EXPECTED TO BE RETURN FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1507816 FARADRIVE STEERABLE SHEATH CLEAR CATHETER, STEERABLE DRA BOSTON SCIENTIFIC CORPORATION CL13974 00191506043131

Patients

Seq Age Sex Outcome Treatment
1 NA Male