UNICEL DXI 800 ACCESS® IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-04080
- Event Type
- Malfunction
- Date Received
- September 25, 2011
- Date of Event
- August 22, 2011
- Report Date
- August 22, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PATIENT SAMPLES WERE COLLECTED IN "ORANGE TOP" AND "GREEN TOP" SAMPLE TUBES. ADDITIONAL COLLECTION DEVICE AND SAMPLE CENTRIFUGATION INFORMATION HAVE NOT BEEN PROVIDED. PER THE CUSTOMER FEEDBACK EVENT DESCRIPTION, QUALITY CONTROL (QC) WAS PERFORMING WITHIN THE CUSTOMER'S ESTABLISHED RANGES AT THE TIME OF THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THE EVENT. THE FSE CLEANED ALL PROBES, WASH STATIONS AND CHECKED SAMPLE PROBE AND ALL REAGENT PIPETTOR ALIGNMENTS. THE FSE REPLACED ALL ASPIRATE PROBES. THE FSE PERFORMED A CARRYOVER TEST AND RESULTS WERE WITHIN INSTRUMENT SPECIFICATIONS.
A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BEC) TO REPORT OBTAINING ERRONEOUSLY ELEVATED TOTAL BETA HUMAN CHORIONIC GONADOTROPIN (T-HCG) RESULTS ABOVE THE NORMAL REFERENCE RANGE FOR THREE (3) PATIENTS GENERATED BY A UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, USED IN CONJUNCTION WITH ACCESS T-HCG REAGENT (LOT 109220) AND CALIBRATOR (LOT 017956). REPEAT TESTING OF THE PATIENTS' SAMPLES PRODUCED LOWER RESULTS WITHIN THE NORMAL REFERENCE RANGE OF THE ASSAY FOR TWO (2) PATIENTS. FOR THE THIRD PATIENT, THE REPEAT RESULTS WERE ELEVATED BUT WERE NOT WITHIN LABELED T-HCG ASSAY PRECISION CLAIMS. ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WERE NO REPORTS OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 800 ACCESS® IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXI 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |