FDA Adverse Event Malfunction Summary report: N

GSM ANTENNA

MDR report key: 22621181 · Received July 25, 2025

Report

Report Number
3004936110-2025-01467
Event Type
Malfunction
Date Received
July 25, 2025
Report Date
July 25, 2025
Manufacturer
ABBOTT MEDICAL
Product Code
MOM
PMA / PMN Number
P100045
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ONE CARDIOMEMS I2 PATIENT ELECTRONICS SYSTEM WAS RECEIVED FOR INVESTIGATION. VISUAL INSPECTION OF RETURNED MATERIAL REVEALED FUNCTIONAL DAMAGE TO THE SMA QUAD BAND 6IN RUBBER DUCK ANTENNA, WHICH IS COMMONLY REFERRED TO AS THE GSM ANTENNA. DAMAGE TO THE GSM ANTENNA CAME IN THE FORM OF THE SMA CONNECTOR PORTION BROKEN OFF FROM THE REST OF THE COMPONENT RESULTING IN EXPOSED ANTENNA CONTACTS. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO MANUFACTURING NON-CONFORMANCE'S WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.

Description of Event or Problem · 0

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A BROKEN GSM ADAPTOR RESULTING IN EXPOSED ANTENNA CONTACTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2639308 GSM ANTENNA System, hemodynamic, implantable MOM ABBOTT MEDICAL CM1040

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown