FDA Adverse Event
Injury
Summary report: N
GMK SPHERE TOTAL KNEE SYSTEM
MDR report key: 22620673
·
Received July 25, 2025
Report
- Report Number
- 3005180920-2025-00716
- Event Type
- Injury
- Date Received
- July 25, 2025
- Date of Event
- July 9, 2025
- Report Date
- July 22, 2025
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JWH
- UDI-DI
- 07630030826535
- PMA / PMN Number
- K121416
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
BATCH REVIEW PERFORMED ON (B)(6) 2025. LOT. 2011125: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21-DEC-2020. EXPIRATION DATE: 01-DEC-2025. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENTS DURING THE PERIOD OF REVIEW. ROOT CAUSE: THE SURGEON REVISED LINER HEIGHT, WHICH IS A COMMON PRACTICE TO RESTORE THE JOINT STABILITY. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT.
Description of Event or Problem · 0
AT ABOUT 3 YEARS 8 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING INSTABILITY AND THE CAUSE IS UNKNOWN. THE SURGEON UPSIZED THE POLY (10 TO 12 MM) TO GIVE THE PATIENT MORE STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1178001 | GMK SPHERE TOTAL KNEE SYSTEM | GMK-SPHERE TIBIAL INSERT - FLEX S4L - 10 MM | JWH | MEDACTA INTERNATIONAL SA | 02.12.0410FL | 2011125 | 07630030826535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Female | Required Intervention |