FDA Adverse Event Malfunction Summary report: N

WARMTOUCH COMPACT WARMING UNIT

MDR report key: 2261737 · Received August 30, 2011

Report

Report Number
2936999-2011-00548
Event Type
Malfunction
Date Received
August 30, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
COVIDIEN/FORMERLY TYCO HEALTHCARE
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND HEAT SETTINGS NOT WORKING, UNIT HEATS CONTINUOUSLY, PCB BOARD WAS DEFECTIVE AND REPLACED. THE MANUFACTURER DATE IS UNK. THE WT-5800 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5300 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND IS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5300 WARMTOUCH IS K020604.

Description of Event or Problem · 1

IT WAS REPORTED TO COVIDIEN THAT UNIT ALWAYS ON SAME TEMPERATURE. THE UNIT GAVE AN ALARM. THE UNIT ALWAYS HEAT ON THE SAME TEMPERATURE, NO STEP DOWN. UNIT IN USE WITH A PT, THERE WAS NO PT HARM REPORTED.