FDA Adverse Event
Malfunction
Summary report: N
WARMTOUCH COMPACT WARMING UNIT
MDR report key: 2261737
·
Received August 30, 2011
Report
- Report Number
- 2936999-2011-00548
- Event Type
- Malfunction
- Date Received
- August 30, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 1, 2011
- Manufacturer
- COVIDIEN/FORMERLY TYCO HEALTHCARE
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION FOUND HEAT SETTINGS NOT WORKING, UNIT HEATS CONTINUOUSLY, PCB BOARD WAS DEFECTIVE AND REPLACED. THE MANUFACTURER DATE IS UNK. THE WT-5800 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5300 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND IS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5300 WARMTOUCH IS K020604.
Description of Event or Problem · 1
IT WAS REPORTED TO COVIDIEN THAT UNIT ALWAYS ON SAME TEMPERATURE. THE UNIT GAVE AN ALARM. THE UNIT ALWAYS HEAT ON THE SAME TEMPERATURE, NO STEP DOWN. UNIT IN USE WITH A PT, THERE WAS NO PT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARMTOUCH COMPACT WARMING UNIT | CONVECTIVE AIR WARMER | DWJ | COVIDIEN/FORMERLY TYCO HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |