FDA Adverse Event Malfunction Summary report: N

UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM

MDR report key: 2261733 · Received September 2, 2011

Report

Report Number
2122870-2011-03286
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
January 10, 2010
Report Date
January 11, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
K023764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2010, THE FIELD SERVICE ENGINEER (FSE) PERFORMED THE HIGH SENSITIVITY SYSTEM CHECK, WHICH FAILED THE FOAM PORTION. FSE DETERMINED THE THREE ASPIRATE PROBES WERE NOT ASPIRATING AT THE SAME RATE AND REPLACED THE PERI PUMP TUBING AND THE ASPIRATE PROBES. HIGH SENSITIVITY SYSTEM CHECK AND A 40 POINT PRECISION RUN MET PUBLISHED SPECIFICATIONS. NO OTHER HARDWARE ISSUES WERE NOTED. ROOT CAUSE WAS NOT DETERMINED, BUT MAY BE RELATED TO THE PARTS REPLACED. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN JANUARY 1 2008 AND OCTOBER 23, 2010 OF COMPLAINTS FOR ADDITIONAL REPORTABLE EVENTS. THIS IS EVENT 5 OF 13 REPORTED BY THIS CUSTOMER. THE 13 MDRS ARE: 212870-2011-03282, 03283, 03284, 03285, 03287, 03288, 03289, 03290, 03291, 03292, 03293 AND 03294.

Description of Event or Problem · 1

CUSTOMER REPORTED ERRONEOUS HIGH ACCU TNI (TROPONIN I) TEST RESULTS WERE OBTAINED FOR 13 PT SAMPLES WHEN USING THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. THE INITIAL TEST RESULTS WERE ABOVE THE NORMAL REFERENCE RANGE, AND BELOW THE AMI (ACUTE MYOCARDIAL INFARCTION) CUTOFF OF 0.50 NG/ML. THE TEST RESULTS WERE REPORTED OUT OF THE LAB. REPEAT ANALYSIS WAS PERFORMED ON ANOTHER UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. THE TEST RESULTS WERE WITHIN THE NORMAL RANGE. IMPACT TO PT TREATMENT IS UNK. NO REPORTS OF DEATH OR SERIOUS INJURY AS A RESULT OF THIS EVENT. THIS IS EVENT 5 OF 13 REPORTED BY THIS CUSTOMER FOR 13 PT SAMPLES TESTED ON (B)(6) 2010.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK ACCESS ACCU TNI