FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2261235 · Received September 23, 2011

Report

Report Number
2531779-2011-07037
Event Type
Injury
Date Received
September 23, 2011
Date of Event
September 19, 2011
Report Date
September 20, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION (B)(6) 2011 DEVICE EVALUATION: A RETAIN CARTRIDGE FROM LOT # 201572 HAS BEEN EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE PATIENT EXPERIENCED A BLOOD GLUCOSE OF 31 MMOL/L. THE FAMILY MEMBER REPORTED THAT THE PATIENT'S BG STARTED ELEVATING ON (B)(6) 2011. THE PUMP REPORTEDLY ALARMED FOR AN EMPTY CARTRIDGE AND THE CARTRIDGE COMPARTMENT WAS WET WITH INSULIN; THE FAMILY MEMBER WAS REPORTEDLY UNSURE IF THE INSULIN WAS COMING FROM THE PLUNGER OR FROM THE LUER CONNECTION. THE CARTRIDGE AND SET WERE CHANGED. ON (B)(6) 2011 DURING A BOLUS, THE FAMILY MEMBER REPORTEDLY NOTICED LEAKING AT THE INFUSION SITE. THE FAMILY MEMBER REPORTED THAT THE SITE WAS REMOVED AND THE CANNULA DID NOT LOOK BENT, THERE WAS NO BLOOD AT THE SITE, AND NO SCAR TISSUE WAS FOUND. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPERGLYCEMIA WHILE USING INSULIN PUMP THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN PUMP CARTRIDGE LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART

Patients

Seq Age Sex Outcome Treatment
1 13 YR Required Intervention