FDA Adverse Event Malfunction Summary report: N

O-ARM O2 IMAGING SYSTEM

MDR report key: 22607879 · Received July 25, 2025

Report

Report Number
3004785967-2025-00513
Event Type
Malfunction
Date Received
July 25, 2025
Date of Event
July 3, 2025
Report Date
September 29, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC. (BOXBOROUGH)
Product Code
OWB
PMA / PMN Number
K200074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(H3,H6) NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. B17,C20,D15 ARE APPLICABLE CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: BI71000429, SERIAL/LOT #: -, UBD: , UDI#: MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THE COMPONENT WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. ANALYSIS FOUND THAT CONFIRM REPORTED COMPLAINT. TEST WINCH MOTOR ASSEMBLY WITH MOTION CART. FORWARD AND BACKWARDS MOTION PASSED, BRAKE WAS ENGAGING AND DISENGAGING AS NORMAL. DRIVE ASSEMBLY FUNCTIONED AS NORMAL. VISUAL INSPECTION SHOWS CABLE STRANDS ARE FRAYING, CABLE SHOWING SIGNS OF WEAR. DAMAGED CABLE. B01,C07,D02 ARE APPLICABLE CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: BI71000429, SERIAL/LOT #: (B)(6) REV. 8 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

(H3,H6): MANUFACTURING REPRESENTATIVE WENT TO THE SITE TO TEST THE SYSTEM, HARDWARE PART WAS REPLACED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING AN IMAGING SYSTEM BEING USED OUTSIDE OF THE PROCEDURE. IT WAS REPORTED THAT THE DOOR COULD NO LONGER OPEN. A MALFUNCTION IN THE DOOR WIRE WAS IDENTIFIED. NO PATIENT WAS PRESENT AT THE TIME OF EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2038604 O-ARM O2 IMAGING SYSTEM INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB MEDTRONIC NAVIGATION, INC. (BOXBOROUGH) BI70002000

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11....".| SEE H11.