RECIPROC BLUE FILES, 6X, STERILE
Report
- Report Number
- 9611053-2025-00108
- Event Type
- Malfunction
- Date Received
- July 25, 2025
- Report Date
- September 5, 2025
- Manufacturer
- VDW GMBH
- Product Code
- EKS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21 CFR PART 803. THIS EVENT WILL BE RE-EVALUATED, AS APPROPRIATE, IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
A RECIPROC BLUE FILE R25 8/100 25MM 025 WAS RETURNED IN LOOSE. THE INSTRUMENT IS BROKEN AT THE TIP OF THE ACTIVE PART (TORQUE). NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1904507). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. FOR INFORMATION, WE REMIND THE PRACTITIONER HAS TO MAKE SURE THAT A STRAIGHT-LINE ACCESS IS CREATED PRIOR USING THE RECIPROC BLUE FILES (AS MENTIONED IN THE DFU). ROOT CAUSES ARE NOT IDENTIFIED. ALL COMPLAINTS ARE MONITORED THROUGH THE MONTHLY PRODUCT SURVEILLANCE COMMITTEE AND A CORRECTIVE ACTION COULD BE DETERMINED BY THE COMMITTEE.
IN THIS EVENT IT IS REPORTED THAT A RECIPROC BLUE FILES, 6X, STERILE FILE BROKE DURING USE. THE OUTCOME OF THIS EVENT IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2038570 | RECIPROC BLUE FILES, 6X, STERILE | FILE, PULP CANAL, ENDODONTIC | EKS | VDW GMBH | 1904507 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |