CAPSTONE® SPINAL SYSTEM
Report
- Report Number
- 1030489-2025-02499
- Event Type
- Malfunction
- Date Received
- July 25, 2025
- Date of Event
- May 12, 2025
- Report Date
- July 25, 2025
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- MQP
- UDI-DI
- 00613994803061
- PMA / PMN Number
- K103731
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H3:PRODUCT ANALYSIS # (B)(4):PART # 2990001, LOT # KH19H114 VISUAL AND OPTICAL INSPECTION CONFIRMED ONE OF THE PRONGS OF THE INSERTER HAS BROKEN. THE DAMAGE TO THE TIP OF THE INSERTER IS CONSISTENT WITH BEND STRESS OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A INSERTER USED FOR SPIN AL THERAPY. IT WAS REPORTED THAT ONE OF THE CLIPS THAT MATES WITH THE IMPLANT WAS MISSING FROM THE BASE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2286004 | CAPSTONE® SPINAL SYSTEM | SPINAL VERTEBRAL BODY REPLACEMENT DEVICE | MQP | MEDTRONIC SOFAMOR DANEK USA, INC | 2990001 | KH19H114 | 00613994803061 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |