FDA Adverse Event Malfunction Summary report: N

RADIOLUCENT DRIVE DRILL BIT

MDR report key: 2260301 · Received September 1, 2011

Report

Report Number
9616696-2011-00143
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 8, 2011
Report Date
August 8, 2011
Manufacturer
STRYKER IRELAND LTD.
Product Code
GFF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL BIT SUBJECT TO THIS INVESTIGATION WAS NOT RETURNED TO THE MANUFACTURER FOR EVAL. THE RRAD (RADIOLUCENT RIGHT ANGLE DRIVE) ATTACHMENT USED WITH THE DRILL BIT WAS RETURNED AND EVALUATED. ON INSPECTION OF THE ATTACHMENT IT WAS CONFIRMED THAT PART OF THE DRILL BIT WAS STILL STUCK IN THE ATTACHMENT. IT WAS VISUALLY CONFIRMED THAT THE DRILL BIT HAD BROKEN FROM THE ULTEM SECTION (DRILL BIT GUIDE) OF THE PRODUCT. LOT NUMBER INFO HAS NOT BEEN PROVIDED TO PERMIT FURTHER INVESTIGATION. THE ROOT CAUSE IS UNDETERMINED. THE ATTACHMENT ASSOCIATED WITH THIS MDR IS AS FOLLOWS: PART NUMBER: 4100355000, SERIAL NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A T2 HUMERUS SURGERY, THE DRILL BIT BROKE AT THE BASE, WHERE THE BIT IS ATTACHED TO THE RRAD ATTACHMENT. IT WAS ALSO REPORTED THAT THERE WAS NO PT OR USER INJURY AND THERE WAS NO DELAY TO SURGERY. IT WAS FURTHER REPORTED THAT ANOTHER DRILL BIT WAS READILY AVAILABLE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RADIOLUCENT DRIVE DRILL BIT GFF STRYKER IRELAND LTD. UNK

Patients

Seq Age Sex Outcome Treatment
1 46 YR