FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2260273 · Received September 22, 2011

Report

Report Number
6000001-2011-26322
Event Type
Malfunction
Date Received
September 22, 2011
Date of Event
July 20, 2011
Report Date
September 5, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK AT THE FILL PORT WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGNS OF PHYSICAL ABNORMALITY. TO VERIFY THE REPORTED DEFECT, THE UNIT WAS MANUALLY FILLED WITH WATER TO ITS MAXIMUM VOLUME. DURING AND AFTER FILL, NO EVIDENCE OF LEAKAGE WAS NOTED AT THE FILL PORT. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR LEAKED AT THE FILL PORT DURING FILLING. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 09J069

Patients

Seq Age Sex Outcome Treatment
1