FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 2260068 · Received August 26, 2011

Report

Report Number
3004209178-2011-06907
Event Type
Injury
Date Received
August 26, 2011
Date of Event
January 1, 2011
Report Date
August 8, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LKK
PMA / PMN Number
P860004
Removal / Correction Number
Z-0592-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2011, IT WAS REPORTED THAT THE PT EXPERIENCED A 30 POUND WEIGHT GAIN IN THE LAST MONTH AND HAD AT LEAST 4 FALLS. TWO OF THOSE FALLS WERE FACE-FORWARD, WHICH LEAD TO BRUISING ON HER CHEST. THE PT WAS NOT SURE IF THE PUMP WAS HIT DURING THE FORWARD FALLS, ONE OF WHICH WAS DOWN A 3-STEP FLIGHT OF STAIRS. DURING A PUMP REFILL THAT OCCURRED THE FIRST WEEK OF (B)(6) 2011, THE PUMP HAD A LARGE VOLUME DISCREPANCY (ACTUAL AMOUNT WAS GREATER THAN EXPECTED). AT THAT TIME, THE PT HAD NO SIGNS/SYMPTOMS OF WITHDRAWAL; HOWEVER ON (B)(6) 2011, THE PT EXPERIENCED INCREASED PAIN AND INCREASED SPASMS. ON (B)(6) 2011, A DYE AND ROTOR STUDY WERE PERFORMED. CSF WAS WITHDRAWN, BUT THE PT HAD A PSEUDOMENINGOCELE, "SO IT MIGHT BE FLUID WHICH ACCUMULATED AT THAT SITE." DURING THE DYE STUDY, THE DYE DID NOT LEAK AND NO FRACTURE WAS SEEN. THE RADIOLOGIST SAW DYE EXIT FROM THE TIP OF THE CATHETER WHICH WAS LOCATED AT L1 (LOWER THAN IMPLANT LEVEL), BUT THE REPORTER DID NOT. THE ROTOR STUDY REVEALED THE MOTOR MOVED COUNTER-CLOCKWISE. THE PLAN WAS TO REFER THE PT TO A SURGEON FOR A REVISION SURGERY. IT WAS LATER REPORTED THAT AN MRI WAS COMPLETED ON (B)(6), AND A MOTOR STALL RECOVERED 35 MINS LATER. PUMP EVENT LOGS WERE OTHERWISE INDICATED AS BEING NORMAL. THE DRUGS INFUSED WERE BACLOFEN 310MCG/ML AT DAILY DOSE OF 87.57 MCG AND DILAUDID 100 MCG/ML.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHROMED II LKK MDT PUERTO RICO OPERATIONS CO., JUNCOS 863740 NA

Patients

Seq Age Sex Outcome Treatment
1 49 YR Female Required Intervention CATHETER: MODEL 8709, LOT# N084798032| CATHETER: MODEL 8709SC, LOT# N123616008| EXPLANTED:| EXPLANTED:| IMPLANTED:| IMPLANTED:| EXPLANTED:| CATHETER: MODEL 8709, LOT # N084798032| IMPLANTED:| CATHETER: MODEL 8709SC, LOT # N123616008| EXPLANTED:| IMPLANTED: