FDA Adverse Event Malfunction Summary report: N

TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE

MDR report key: 22593524 · Received July 24, 2025

Report

Report Number
9610773-2025-04703
Event Type
Malfunction
Date Received
July 24, 2025
Date of Event
June 26, 2025
Report Date
October 23, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
EWY
UDI-DI
04042761052764
PMA / PMN Number
K950076
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, G1, G3, H2, H3, H6, H11. THE DEVICE WAS RETURNED TO REGIONAL SERVICE CENTER FOR INSPECTION, AND THE REPORTED FAILURE WAS CONFIRMED. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: A COMPONENT FAILURE OF THE LENS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE HAD LENS COMES OFF. THE ISSUE WAS IDENTIFIED DURING DEVICE INSPECTION BEFORE USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1343791 TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE RIGID OPTICAL LAPAROSCOPE EWY OLYMPUS WINTER & IBE GMBH WA53005A 04042761052764

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown