FDA Adverse Event Malfunction Summary report: N

TRIEVER24

MDR report key: 22588264 · Received July 24, 2025

Report

Report Number
3020347218-2025-00047
Event Type
Malfunction
Date Received
July 24, 2025
Date of Event
June 26, 2025
Report Date
July 24, 2025
Manufacturer
INARI MEDICAL, INC.
Product Code
QEW
UDI-DI
0081023710096
PMA / PMN Number
K213402
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INARI DEVICES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. THE DEVICE IDENTIFIERS WERE PROVIDED, AND THE LOT HISTORY RECORDS WERE REVIEWED. THERE WERE NO DISCREPANCIES OR UNUSUAL FINDINGS. ADDITIONALLY, A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND NO SIMILAR ISSUES HAVE BEEN REPORTED AGAINST THIS LOT NUMBER. THE DEVICE LABELING INCLUDES THE FOLLOWING WARNING STATEMENTS: "AVOID USING EXCESSIVE FORCE TO ADVANCE OR RETRACT THE FLOWTRIEVER CATHETER OR TRIEVER24 AGAINST RESISTANCE. IF EXCESSIVE RESISTANCE OCCURS, RETRACT, AND REMOVE THE DEVICE. EXCESSIVE FORCE AGAINST RESISTANCE MAY RESULT IN DAMAGE TO THE DEVICE OR VESSEL PERFORATION." MANUFACTURER REFERENCE: (B)(4).

Description of Event or Problem · 0

ON (B)(6) 2025 A 73-YEAR-OLD FEMALE PATIENT UNDERWENT PULMONARY EMBOLISM (PE) THROMBECTOMY USING INARI DEVICES. AFTER THE THIRD ASPIRATION USING THE TRIEVER24 CATHETER (T24), THE PHYSICIAN NOTICED THE DEVICE WAS OCCLUDED AND NO CLOT WAS REMOVED. NO KINKS IN THE CATHETER WERE OBSERVED. THE PHYSICIAN ATTEMPTED TO REINSERT THE DILATOR BUT WAS MET WITH SIGNIFICANT RESISTANCE. THE CATHETER LINER HAD SEPARATED AND CAME OUT OF THE DEVICE WHEN THE CATHETER WAS FLUSHED. A NEW DEVICE WAS OPENED, AND THE CASE WAS COMPLETED WITHOUT FURTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2188518 TRIEVER24 PERIPHERAL MECHANICAL THROMBECTOMY WITH ASPIRATION QEW INARI MEDICAL, INC. 22-101 25010147 0081023710096

Patients

Seq Age Sex Outcome Treatment
1 73 YR Female FLOWSAVER| FLOWTRIEVER XL CATHETER| INTRI24 INTRODUCER SHEATH| TRIEVER16 CATHETER