FDA Adverse Event Malfunction Summary report: N

BD BBL¿ CHROMAGAR¿ ORIENT/TSA II W/5% SB

MDR report key: 22584648 · Received July 24, 2025

Report

Report Number
1119779-2025-01882
Event Type
Malfunction
Date Received
July 24, 2025
Date of Event
January 19, 2024
Report Date
June 12, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JSH
UDI-DI
10382902222399
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

"INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 222239, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 3320027. NO RETENTION SAMPLES FOR THIS BATCH WERE AVAILABLE FOR INVESTIGATION. NO RETURN SAMPLES WERE RECEIVED FOR INVESTIGATION OF THIS COMPLAINT. FIVE PHOTOS WERE RECEIVED FOR INVESTIGATION OF THIS COMPLAINT. PHOTOS SHOW CONTAMINATION IN BATCH 3320027 WITH TIMESTAMP 1510 AND 1608 VISIBLE IN SOME OF THE PHOTOS. THIS COMPLAINT CAN BE CONFIRMED. NO COMPLAINT TRENDS HAVE BEEN IDENTIFIED ON THIS PRODUCT FOR THIS DEFECT. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION." THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING APRIL 2023-2025, UNDER CAPA 11910483.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD BBL¿ CHROMAGAR¿ ORIENT/TSA II W/5% SB, THERE WAS A FALSE RESULT DUE TO CONTAMINATION. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1591648 BD BBL¿ CHROMAGAR¿ ORIENT/TSA II W/5% SB CULTURE MEDIA, NON-SELECTIVE AND DIFFERENTIAL JSH BECTON DICKINSON & CO. (SPARKS) 3320027 10382902222399

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown