FDA Adverse Event Summary report: N

4085 SURGICAL TABLE

MDR report key: 2258392 · Received September 21, 2011

Report

Report Number
1043572-2011-00043
Date Received
September 21, 2011
Date of Event
August 30, 2011
Report Date
September 21, 2011
Manufacturer
STERIS CORPORATION - MONTGOMERY
Product Code
FQO
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE STERIS ACCOUNT MANAGER AND SERVICE TECHNICIAN INSPECTED THE TABLE AND FOUND THAT THE SHROUD WHICH HAD BEEN REMOVED BY FACILITY PERSONNEL WAS DAMAGED. SPECIFICALLY, THE TWO BOTTOM SECTIONS OF THE SHROUD WERE BENT AT THE BOTTOM EDGE, WHICH CAUSED THE SECTIONS TO BIND WHEN THE TABLE WAS COMMANDED TO MOVE DURING THE REPORTED EVENT. THE TECHNICIAN REPLACED THE SHROUDS, TESTED AND RETURNED THE TABLE TO USE. NO FURTHER ISSUES HAVE BEEN REPORTED WITH THE SURGICAL TABLE SINCE THE REPORTED EVENT. DURING THEIR INVESTIGATION OF THE EVENT, STERIS PERSONNEL IDENTIFIED THAT THE USER FACILITY HAS BEEN USING THE BASE OF THEIR SURGICAL TABLES FOR STORAGE. STORAGE OF ITEMS ON THE TABLE BASE CAN CAUSE BENDING OF THE SHROUD AS EVIDENCED IN THIS EVENT AND IS CONTRAINDICATED IN THE LABELING OF THE TABLE. THE TABLE'S WARNINGS INCLUDE A LABEL ON THE BASE THAT STATES " DO NOT USE TABLE BASE FOR STORAGE." THE OPERATOR MANUAL (PP. 3-1) STATES: " WARNING- PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD: DO NOT STORE ITEMS ON TABLE BASE. DOING SO MAY RESULT IN EQUIPMENT DAMAGE OR INADVERTENT TABLETOP MOVEMENT PLACING THE PATIENT AND/OR USER AT RISK OF INJURY." IN-SERVICE TRAINING ON THE PROPER USE AND OPERATION OF THE SURGICAL TABLE HAS BEEN SCHEDULED FOR (B)(6) 2011.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT DURING A SURGICAL PROCEDURE, HOSPITAL STAFF COMMANDED THE TABLE TO MOVE INTO A TRENDELENBURG POSITION AND HEARD A BINDING SOUND. THE STAFF OBSERVED THAT THE TELESCOPING SHROUD ON THE TABLE EVIDENCED DAMAGE. THEY HAD THE SHROUD REMOVED, POSITIONED THE TABLE AND COMPLETED THE PROCEDURE SUCCESSFULLY. THERE WAS NO INJURY TO HOSPITAL STAFF OR PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4085 SURGICAL TABLE OPERATING TABLE FQO STERIS CORPORATION - MONTGOMERY 4085

Patients

Seq Age Sex Outcome Treatment
1 Other