PERCLOSE¿ PROSTYLE¿
Report
- Report Number
- 2024168-2025-07653
- Event Type
- Injury
- Date Received
- July 23, 2025
- Date of Event
- July 2, 2025
- Report Date
- September 23, 2025
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. E1: (B)(6).
CORRECTIONS: D4 - LOT #: UPDATED FROM 5041842 TO 5042241. D4 - PRIMARY UDI NUMBER: UPDATED FROM (B)(4). H4 - DEVICE MFG DATE: UPDATED FROM 4/18/2025 TO 4/22/2025. E1: (B)(6).
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 7F SHEATH HOLE PRIOR TO AN INTERVENTIONAL TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) PROCEDURE. REPORTEDLY, WITH THE PROSTYLE DEVICE, THE SUTURE WAS NOT ATTACHED TO THE NEEDLE WHEN THE PLUNGER WAS REMOVED. THE SUTURE OF A NEW PROSTYLE DEVICE WAS SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 16F SHEATH, AND THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURE. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 7F SHEATH HOLE PRIOR TO AN INTERVENTIONAL (TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) PROCEDURE. REPORTEDLY, WITH THE PROSTYLE DEVICE, THE SUTURE WAS NOT ATTACHED TO THE NEEDLE WHEN THE PLUNGER WAS REMOVED. THE SUTURE OF A NEW PROSTYLE DEVICE WAS SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 16F SHEATH, AND THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURE. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1774640 | PERCLOSE¿ PROSTYLE¿ | DEVICE, HEMOSTASIS, VASCULAR | MGB | ABBOTT VASCULAR INC. | 12773-03 | 5042241 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |