EVIS EXERA III COLONOVIDEOSCOPE
Report
- Report Number
- 9610595-2025-15165
- Event Type
- Malfunction
- Date Received
- July 23, 2025
- Date of Event
- July 2, 2025
- Report Date
- July 23, 2025
- Manufacturer
- AIZU OLYMPUS CO., LTD.
- Product Code
- FDF
- UDI-DI
- 04953170305115
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. BASED ON THE RESULTS OF THE INVESTIGATION, THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. THE SUBJECT EVENT CAN BE DETECTED AND PREVENTED BY IMPLEMENTING IN ACCORDANCE WITH THE BELOW INSTRUCTIONS FOR USE (IFU): NOTHING OTHER THAN STERILE WATER SHOULD BE USED FOR AIR/WATER FEEDING. NO ADDITIVES SHOULD BE PUT INTO THE STERILE WATER. NON-STERILE WATER MAY CAUSE PATIENT CROSS-CONTAMINATION AND/OR INFECTION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE AIR/WATER TUBE (A/W-TUBE), JET TUBE, AND THE AIR/WATER CYLINDER (A/W-CYLINDER). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 986092 | EVIS EXERA III COLONOVIDEOSCOPE | COLONOVIDEOSCOPE | FDF | AIZU OLYMPUS CO., LTD. | CF-HQ190L | 04953170305115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |