FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III COLONOVIDEOSCOPE

MDR report key: 22580949 · Received July 23, 2025

Report

Report Number
9610595-2025-15165
Event Type
Malfunction
Date Received
July 23, 2025
Date of Event
July 2, 2025
Report Date
July 23, 2025
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDF
UDI-DI
04953170305115
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. BASED ON THE RESULTS OF THE INVESTIGATION, THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. THE SUBJECT EVENT CAN BE DETECTED AND PREVENTED BY IMPLEMENTING IN ACCORDANCE WITH THE BELOW INSTRUCTIONS FOR USE (IFU): NOTHING OTHER THAN STERILE WATER SHOULD BE USED FOR AIR/WATER FEEDING. NO ADDITIVES SHOULD BE PUT INTO THE STERILE WATER. NON-STERILE WATER MAY CAUSE PATIENT CROSS-CONTAMINATION AND/OR INFECTION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE AIR/WATER TUBE (A/W-TUBE), JET TUBE, AND THE AIR/WATER CYLINDER (A/W-CYLINDER). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
986092 EVIS EXERA III COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF AIZU OLYMPUS CO., LTD. CF-HQ190L 04953170305115

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown