FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 22580940 · Received July 23, 2025

Report

Report Number
2955842-2025-31326
Event Type
Malfunction
Date Received
July 23, 2025
Date of Event
July 2, 2025
Report Date
July 2, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112298
PMA / PMN Number
K220023
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WAS ANALYZED AND FOUND TO HAVE BLADE DAMAGE AT THE TIP OF THE BLADE. BOTH OF THE BLADE EDGES WAS INDENTED. THE BLADE INDENTATION DID CAUSE THE CUT TEST FAILURE AND CAUSED THE BLADES NOT TO OPEN OR CLOSE PROPERLY. THE CUTTING EDGE HAD CORROSION THAT WOULD HAVE CONTRIBUTED TO THE BLADE DAMAGE OR CUTTING FAILURE. THE COMPLAINT REGARDING GOUGE IN SCISSOR BLADE WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF NICKS AND GOUGES ON THE INSTRUMENT BLADES IS ATTRIBUTED TO USE-RELATED DAMAGE WHEN ATTEMPTING TO CUT INCOMPATIBLE MATERIAL (I.E., HARD OBJECTS SUCH AS BONE OR CARTILAGE) OR MISHANDLING THE INSTRUMENT.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL INC. (ISI), HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT HAD A GOUGE IN THE BLADE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
986083 ENDOWRIST MONOPOLAR CURVED SCISSORS NAY INTUITIVE SURGICAL, INC 470179-23 K10250115 0166 00886874112298

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES