ENDOWRIST
Report
- Report Number
- 2955842-2025-31326
- Event Type
- Malfunction
- Date Received
- July 23, 2025
- Date of Event
- July 2, 2025
- Report Date
- July 2, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874112298
- PMA / PMN Number
- K220023
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WAS ANALYZED AND FOUND TO HAVE BLADE DAMAGE AT THE TIP OF THE BLADE. BOTH OF THE BLADE EDGES WAS INDENTED. THE BLADE INDENTATION DID CAUSE THE CUT TEST FAILURE AND CAUSED THE BLADES NOT TO OPEN OR CLOSE PROPERLY. THE CUTTING EDGE HAD CORROSION THAT WOULD HAVE CONTRIBUTED TO THE BLADE DAMAGE OR CUTTING FAILURE. THE COMPLAINT REGARDING GOUGE IN SCISSOR BLADE WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF NICKS AND GOUGES ON THE INSTRUMENT BLADES IS ATTRIBUTED TO USE-RELATED DAMAGE WHEN ATTEMPTING TO CUT INCOMPATIBLE MATERIAL (I.E., HARD OBJECTS SUCH AS BONE OR CARTILAGE) OR MISHANDLING THE INSTRUMENT.
INTUITIVE SURGICAL INC. (ISI), HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT HAD A GOUGE IN THE BLADE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 986083 | ENDOWRIST | MONOPOLAR CURVED SCISSORS | NAY | INTUITIVE SURGICAL, INC | 470179-23 | K10250115 0166 | 00886874112298 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |