MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2025-07919
- Event Type
- Injury
- Date Received
- July 23, 2025
- Date of Event
- March 31, 2025
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000303
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RIGHT CAPSULAR CONTRACTURE; BAKER GRADE IV. D4: UDI: AS THE LOT, SERIAL NUMBER, AND CORRESPONDING EXPIRATION DATE FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON AUGUST 11, 2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. PER DEVICE RECEIVED, THE FOLLOWING INFORMATION HAS BEEN UPDATED ON THIS FORM: - LOT NUMBER UNDER FIELD D4 HAS BEEN UPDATED TO "9971103" D4: UDI: AS THE SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. - REPLACEMENT DEVICE USED WAS 3503001BC; SERIAL NUMBER (B)(6). ON AUGUST 18, 2025, DEVICE EVALUATION WAS COMPLETED AS FOLLOWS: MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. DURING VISUAL EVALUATION NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE RETURNED DEVICE. REGARDING THE REPORTED CAPSULAR CONTRACTURE CONDITION, THERE ARE NOT ENOUGH DETAILS TO DETERMINE THE FACTORS THAT MAY HAVE CAUSED THIS CONDITION. CAPSULAR CONTRACTURE IN THE PATIENT¿S BREAST MIGHT BE THE RESULT OF THE BODY¿S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. FURTHER, CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH 350CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE; BAKER GRADE IV ON HER RIGHT SIDE POSTOPERATIVELY. AS A RESULT, THE DEVICE WAS EXPLANTED ON (B)(6) 2025.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1869864 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 9971103 | 00081317000303 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Female | Required Intervention |