FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 22573441 · Received July 23, 2025

Report

Report Number
2955842-2025-31196
Event Type
Malfunction
Date Received
July 23, 2025
Date of Event
June 27, 2025
Report Date
June 27, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL INC., (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) ERBE TO RESOLVE THE REPORTED ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI HAS RECEIVED THE UNIT; HOWEVER, FAILURE ANALYSIS INVESTIGATION ARE STILL ONGOING. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE ERBE UNIT WAS ANALYZED, AND FAILURE ANALYSIS INVESTIGATIONS REPLICATED AND CONFIRMED THE CUSTOMER-REPORTED COMPLAINT. THE REPORTED ISSUE WAS CONFIRMED IN THE FIELD AS ERROR 25913: M-02 MODULE TIMEOUT. VISUAL INSPECTION REVEALED STAIN ON THE FRONT BEZEL. DURING STARTUP, THE ERBE UNIT DISPLAYED ERRORS M-02-3,4-71 AND FAILED THE INSTRUMENT DETECTION TEST ON ALL SOCKETS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION AND FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF ERROR M-02 IS ATTRIBUTED TO A FAULTY COMPONENT OF THE ERBE GENERATOR. M-02 ERRORS INDICATE A MODULE TIMEOUT ISSUE AND, THEREFORE, THE ACTIVATION WAS INTERRUPTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI ASSISTED SURGICAL PROCEDURE, NONE OF THE MONOPOLAR OR BIPOLAR CORDS WERE WORKING. AN INTEGRATED ELECTROSURGICAL UNIT (IESU) ERBE WAS FAULTING WITH ERROR M-02. AN INTUITIVE SURGICAL INC., (ISI) TECHNICAL SUPPORT ENGINEER (TSE) WALKED CUSTOMER THROUGH POWER CYCLE OF ERBE. CUSTOMER WAS UNABLE TO CONFIRM IF ENERGY WAS WORKING AS THEY WERE NOT USING ENERGY INSTRUMENTS AT THE TIME OF CALL. HOWEVER, THE GENERATOR SCREEN LOOKED LIKE IT WILL WORK. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY. AN ISI FIELD SERVICE ENGINEER (FSE) TO FOLLOW-UP.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2724749 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-19 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES