FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 22573434 · Received July 23, 2025

Report

Report Number
2955842-2025-31199
Event Type
Malfunction
Date Received
July 23, 2025
Date of Event
June 26, 2025
Report Date
January 20, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID NOT RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE INTEGRATED ELECTROSURGICAL UNIT (IESU) OR ERBE WAS ANALYZED AND FOUND THE REPORTED C-34 ERROR WAS CONFIRMED AND REPRODUCED DURING POWER-ON TESTING, INDICATING THE FAULT OCCURRED IN THE FIELD. A VISUAL INSPECTION REVEALED NO RELATED ISSUES. THE UNIT WILL BE SENT TO THE ORIGINAL EQUIPMENT MANUFACTURER FOR FURTHER ANALYSIS. THE PROBABLE CAUSE OF ERROR C-34 IS ATTRIBUTED TO A FAULTY ELECTRICAL COMPONENT INSIDE THE ERBE GENERATOR. THIS ERROR INDICATES A LOWER VOLTAGE THAN EXPECTED DURING ENERGY ACTIVATION. THIS ERROR CAN BE RESOLVED THROUGH TROUBLESHOOTING OR REPLACEMENT OF THE GENERATOR.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECTED DEVICE EXHIBITED ENDOSCOPY IMAGES THAT COULD NOT BE DISPLAYED AND PRINTED CIRCUIT BOARD FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2785330 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-29 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES