AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTER
Report
- Report Number
- 9611451-2025-00671
- Event Type
- Malfunction
- Date Received
- July 23, 2025
- Date of Event
- May 2, 2025
- Report Date
- July 23, 2025
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- CAF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER
Narratives
(B)(6). 900PT562 IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO THE 900PT563 WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K211560. PRODUCT BACKGROUND: THE HEATED BREATHING TUBE (HBT) IS A COMPONENT DESIGNED FOR USE WITH THE AIRVO 2 HUMIDIFIER (AIRVO 2), FOR THE DELIVERY OF HUMIDIFIED RESPIRATORY GASES TO PATIENTS, INCLUDING THOSE WHO ARE RECEIVING NASAL HIGH FLOW (NHF) THERAPY. NHF THERAPY IS INTENDED FOR USE WITH SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH FLOW WARMED AND HUMIDIFIED RESPIRATORY GASES. THE AIRVO 2 DEVICE SHOULD NOT BE USED FOR LIFE SUPPORT PURPOSES, AND APPROPRIATE PATIENT MONITORING MUST BE USED AT ALL TIMES. THE HBT AS PART OF THE AIRVO 2 SYSTEM CONTAINS A HEATER WIRE ENCAPSULATED IN PLASTIC WHICH ENSURES OPTIMAL TEMPERATURE AND HUMIDIFICATION LEVELS ARE DELIVERED TO THE PATIENT INTERFACE WHILE MINIMIZING THE AMOUNT OF CONDENSATE IN THE TUBE. METHOD: THE SUBJECT HBT AS PART OF THE 900PT562 AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE (F&P) FOR EVALUATION. OUR INVESTIGATION IS BASED ON THE PHOTOGRAPHS AND DESCRIPTION OF EVENTS PROVIDED BY THE CUSTOMER, PREVIOUS INVESTIGATIONS OF SIMILAR COMPLAINTS AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: VISUAL INSPECTION OF THE PROVIDED PHOTOGRAPH REVEALED THAT A SECTION OF THE HBT WAS SOFTENED AND MELTED. MARKINGS OF THE MELTED HBT WERE ALSO SEEN ON THE BEDDING. CONCLUSION: WE ARE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT. HOWEVER, DAMAGE TO THE HBT MAY BE CAUSED BY FACTORS SUCH AS INCORRECT SET-UP OR COVERING IT WITH A MATERIAL OR OBJECT FOR A PROLONGED PERIOD. VISUAL INSPECTION OF THE PROVIDED PHOTOGRAPH REVEALED THAT MARKINGS OF THE MELTED HBT WERE SEEN ON THE BEDDING. THE 900PT562 AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR USER INSTRUCTIONS SHOW IN PICTORIAL FORMAT THE CORRECT PLACEMENT OF THE DEVICE AND INCLUDES THE FOLLOWING INFORMATION: - "CONNECT BREATHING TUBE CLIP TO PATIENT CLOTHING OR BEDDING." - "NEVER OPERATE THE UNIT IF THE BREATHING TUBE HAS BEEN DAMAGED WITH HOLES, TEARS OR KINKS" - "DO NOT BLOCK THE FLOW OF AIR THROUGH THE UNIT AND BREATHING TUBE." - "DO NOT ADD HEAT TO ANY PART OF THE BREATHING TUBE E.G., COVERING WITH A BLANKET, OR HEATING IT IN AN INCUBATOR OR OVERHEAD HEATER FOR A NEONATE, AS THIS COULD RESULT IN SERIOUS INJURY." - "DO NOT ALLOW THE BREATHING TUBE TO REMAIN IN DIRECT CONTACT WITH SKIN FOR PROLONGED PERIODS OF TIME." ALL HBTS AS PART OF THE AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR ARE VISUALLY INSPECTED AND UNDERGO FUNCTIONAL TESTS, INCLUDING SOAK AND TEMPERATURE, AND HEATER WIRE RESISTANCE. THE HEATED BREATHING TUBES ARE 100% VISUALLY INSPECTED USING A CAMERA SYSTEM. THE HBTS ARE ALSO TESTED FOR RESISTANCE, CONTINUITY, POLARITY AND PITCH DURING PRODUCTION. ADDITIONALLY, A FUNCTIONAL TEST IS CONDUCTED UNDER LOAD. THE HBT WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT THE TIME OF PRODUCTION. THE AIRVO 2 SYSTEM IS DESIGNED TO COMPLY WITH THE ELECTRICAL SAFETY STANDARD IEC 60601-1: 2005+A1:2012. THE CASE IS COMPOSED OF A FLAME RETARDANT MATERIAL. THE SURFACE TEMPERATURE OF THE HBT, WHEN USED IN ACCORDANCE WITH USER INSTRUCTIONS, IS DESIGNED TO BE WITHIN THE LIMITS SPECIFIED BY ISO 8185 WITH REGARD TO HOT TUBE SURFACE TEMPERATURE NOT EXCEEDING 44° CELSIUS. IT IS AN INHERENT RISK OF HBTS, THAT ADDITIONAL HEAT (ABOVE AMBIENT LEVELS) ADDED TO ANY PART OF THE TUBE VIA AN EXTERNAL SOURCE OR BEING COVERED WITH MATERIAL, MAY LEAD TO THE TUBING BECOMING DAMAGED. TO ADDRESS THIS INHERENT RISK AND AS IS REQUIRED UNDER ISO 80601, THE USER INSTRUCTIONS FOR THE 900PT562 CONTAINS THE WARNING "DO NOT ADD HEAT TO ANY PART OF THE BREATHING TUBE E.G. COVERING WITH A BLANKET, OR HEATING IT IN AN INCUBATOR OR OVERHEAD HEATER FOR A NEONATE, AS THIS COULD RESULT IN SERIOUS INJURY." ADDITIONAL "DO NOT COVER" TAGS ARE ALSO ATTACHED TO ALL HBTS TO ALERT THE USER THAT THE HBT SHOULD NOT BE COVERED. THERE ARE MANY SAFETY FEATURES INCORPORATED INTO THE AIRVO 2 SYSTEM TO PREVENT OVERHEATING AND FIRE. THESE INCLUDE: - THE HEATER WIRES IN THE HBT ARE COMPLETELY INSULATED FROM THE GAS PATH. - THE PCB AT THE [PATIENT] END OF THE HBT IS OVER MOULDED WITH THE THERMOPLASTIC POLYMER POLYPROPYLENE, ENSURING IT IS EXCLUDED FROM THE GAS PATH. - THE AIRVO 2 DEVICE CONTAINS TECHNOLOGY WHICH DETECTS SHORT CIRCUITS AND TRANSIENT CURRENT EVENTS IN THE HBT. WHEN DETECTED, THE AIRVO REMOVES POWER TO THE HBT. THE AIRVO 2 PERFORMS THIS DETECTION AT ANY TIME IT IS TURNED ON AND CONNECTED TO THE HBT. THIS FUNCTIONALITY IS CHECKED BY THE CONTROL SYSTEM EACH TIME THE AIRVO IS POWERED UP, OR WHEN A NEW HBT IS CONNECTED. - AN 'OVER-TEMPERATURE' SENSOR WILL AUTOMATICALLY CUT POWER TO THE MOTOR, HEATER PLATE AND HEATER WIRE IF IT DETECTS ANY OVERHEATING AT THE CHAMBER OR THE PATIENT END OF THE HBT. - THE AIRVO 2 DEVICE IS CONTINUOUSLY CHECKS POWER IN THE HBT AND DISABLES THE HEATER WIRE IF THE MEASURED POWER IS TOO HIGH
A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN SOUTH AFRICA THAT THE HEATED BREATHING TUBE AS PART OF THE 900PT562 AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR WAS FOUND MELTED. THE HEALTHCARE FACILITY REPORTED THAT THE SUBJECT DEVICE WAS DISCONNECTED FROM THE PATIENT AT THE TIME OF THE EVENT. THE HEALTHCARE FACILITY FURTHER REPORTED THAT THE SUBJECT DEVICE HAS BEEN DISCARDED. THERE WERE NO PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1937500 | AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTER | HEATED BREATHING TUBE | CAF | FISHER & PAYKEL HEALTHCARE LTD | 900PT562 | NOT PROVIDED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |