FDA Adverse Event
Injury
Summary report: N
*
MDR report key: 225687
·
Received June 1, 1999
Report
- Report Number
- 225687
- Event Type
- Injury
- Date Received
- June 1, 1999
- Date of Event
- May 28, 1999
- Report Date
- June 1, 1999
- Manufacturer
- VASCOR MEDICAL
- Product Code
- DTB
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
VASCOR MODEL 111-256, BIPOLAR IMPEDANCE AT IMPLANT, MEASURED 510 OHMS ON 11/12/96. BIPOLAR IMPEDANCE SLOWLY CLIMBED AS FOLLOWS: 12/3/96 - 640 OHMS; 12/31/96 - 739 OHMS; 2/11/97 - 785 OHMS; 5/20/97 - 945 OHMS. IMPEDANCE HAS SLOWLY BEEN DECLINING SINCE REACHING A HIGH OF 945 OHMS. RANGED FROM 914 OHMS TO 843 OHMS 2/12/99, AND 811-836 OHMS FROM 8/11/97 - 2/12/99. IMPEDANCES BIPOLAR ON 5/28/99 MEASURED 778-826 OHMS. PT IS TOTALLY PACEMAKER DEPENDENT. WILL REPLACE THE LEAD PRIOR TO FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | * Implant | VENTRICULAR PACING LEAD | DTB | VASCOR MEDICAL | 111-256 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Life Threatening |