FDA Adverse Event Malfunction Summary report: N

CLARISSE ONE STEP HCG PREGNANCY TEST

MDR report key: 22563469 · Received July 22, 2025

Report

Report Number
3014171027-2025-00001
Event Type
Malfunction
Date Received
July 22, 2025
Date of Event
June 10, 2025
Report Date
June 11, 2025
Manufacturer
WEIHAI KANGZHOU BIOTECHNOLOGY ENGINEERING CO.,LTD
Product Code
LCX
UDI-DI
00808829125672
PMA / PMN Number
K183097
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE RETAINED SAMPLES FROM THE COMPLAINED BATCH WITH LOT NUMBER #241110 HAVE BEEN EXAMINED. THE DEVICE PERFORMED ACCORDING TO SPECIFICATIONS AND DESIGN INTENT. PRODUCTION RECORDS FOR THIS BATCH HAVE ALSO BEEN REVIEWED. NO NONCONFORMITIES WERE FOUND. ADDITIONALLY, IT WAS CHECKED WHETHER THERE WERE ANY OTHER CUSTOMER COMPLAINTS FOR THIS BATCH, AND IT WAS CONFIRMED THAT NO OTHER COMPLAINTS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

I HAVE USED DOLLAR TREE PREGNANCY TESTS ON AND OFF FOR ABOUT 3 YEARS AND NEVER HAD ANY ISSUES UNTIL NOW. RECENTLY I BOUGHT ONE AND IT COMPLETELY FAILED LIKE NO CONTROL LINE, TOTALLY UNUSABLE. FIGURED IT MIGHT HAVE BEEN A DUD, SO I WENT BACK AND GRABBED ANOTHER. SAME THING HAPPENED AGAIN. I DECIDED TO CALL THE NUMBER ON THE PACKAGE (WWW.DELTABRANDS.COM)TO REPORT IT, JUST TO LET THEM KNOW. THE GUY WHO ANSWERED WAS SUPER DISMISSIVE RIGHT FROM THE START AND SAID,"DID NOT YOU CALL YESTERDAY?''. I EXPLAINED THE SITUATION AND MENTIONED THE SECOND TEST ALSO DID NOT WORK. HE TOLD ME "SHOULD NOT BE BUYING MORE " AND LITERALLY HUNG UP ONE ME. IT WAS SO UNPROFESSIONAL AND HONESTLY MADE ME FEEL STUPID FOR EVEN BOTHERING TO REACH OUT . I AM NOT ABOUT TO WALK INTO DOLLAR TREE WITH A USED TEST IN HAND JUST TO PROVE A POINT , BUT I HAVE RELIED ONE THESE FOR YEARS BECAUSE THEY WERE ALWAYS AFFORDABLE AND ACCURATE . WITH HOW HARD IT IS TO GET A DOCTOR'S APPOINTMENT QUICKLY, THESE TESTS HAVE BEEN A GO-TO FOR ME. I AM CURRENTLY KEEPING A CLOSE EYE ON THINGS AND GETTING BLOOD DRAWN AT (B)(6) JUST TO BE SAFE, BUT IT REALLY SUCKS TO FEEL DISMISSED WHEN YOU ARE JUST TRYING TO TAKE CARE OF YOUR HEALTH. I HAVE ALREADY CONTACTED DOLLAR TREE'S CUSTOMER SUPPORT AS WELL AS THE STORE REPRESENTATIVE I SPOKE WITH WAS DISMISSIVE AND ENDED THE PRODUCT PERFORMANCE OF PROVIDE A RESOLUTION. BECAUSE OF THIS , I AM NOW SEEKING ALTERNATIVE SUPPORT OF GUIDANCE,AS I WENT TO ENSURE MY CONCERNS ARE ACKNOWLEDGED AND ADDRESSED APPROPRIATELY . PATIENT CODE : 4582 . DEVICE CODE : 2588,1506. REFERENCE REPORT : MW5171742.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1784258 CLARISSE ONE STEP HCG PREGNANCY TEST KIT , TEST , PREGNANCY ,HCG,OVER THE COUNTER LCX WEIHAI KANGZHOU BIOTECHNOLOGY ENGINEERING CO.,LTD N/A 241110 00808829125672

Patients

Seq Age Sex Outcome Treatment
1 24 YR Female