FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

MDR report key: 22561508 · Received July 21, 2025

Report

Report Number
9610595-2025-14739
Event Type
Malfunction
Date Received
July 21, 2025
Date of Event
June 24, 2025
Report Date
July 21, 2025
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. THE EVENT CAN BE DETECTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: "AW-CYLINDER (TUBE) HAS FOREIGN OBJECTS" AND "AW-TUBE HAS FOREIGN OBJECTS." THE WHITE FOREIGN MATERIAL DETECTED BY INCOME INSPECTION HAS BEEN HANDLED BY OMSC CAPA-201632. ACCORDING TO PREVIOUS INVESTIGATIONS, THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED AW-CYLINDER (TUBE) HAD FOREIGN OBJECTS AND DEFORMATION OF PLUG UNIT. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
767612 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE GASTROINTESTINAL VIDEOSCOPE FDS AIZU OLYMPUS CO., LTD. GIF-H185

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown