EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Report
- Report Number
- 9610595-2025-14739
- Event Type
- Malfunction
- Date Received
- July 21, 2025
- Date of Event
- June 24, 2025
- Report Date
- July 21, 2025
- Manufacturer
- AIZU OLYMPUS CO., LTD.
- Product Code
- FDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. THE EVENT CAN BE DETECTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: "AW-CYLINDER (TUBE) HAS FOREIGN OBJECTS" AND "AW-TUBE HAS FOREIGN OBJECTS." THE WHITE FOREIGN MATERIAL DETECTED BY INCOME INSPECTION HAS BEEN HANDLED BY OMSC CAPA-201632. ACCORDING TO PREVIOUS INVESTIGATIONS, THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED AW-CYLINDER (TUBE) HAD FOREIGN OBJECTS AND DEFORMATION OF PLUG UNIT. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 767612 | EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE | GASTROINTESTINAL VIDEOSCOPE | FDS | AIZU OLYMPUS CO., LTD. | GIF-H185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |