FDA Adverse Event Malfunction Summary report: N

SFXSYM PDS+ UNI VIO 18IN 1 SA CT

MDR report key: 22559747 · Received July 21, 2025

Report

Report Number
2210968-2025-08429
Event Type
Malfunction
Date Received
July 21, 2025
Date of Event
June 23, 2025
Report Date
August 26, 2025
Manufacturer
ETHICON INC.
Product Code
NEW
PMA / PMN Number
K141776
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4) THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3 H3 INVESTIGATION SUMMARY: A FAILED SUTURE NEEDLE, LABELED AS (B)(4), WAS SUBMITTED TO HERRERA, STAFFORD AND ASSOCIATES, LLC (HSA) FOR FRACTOGRAPHIC EVALUATION ON JULY 11, 2025. THE COMPONENT WAS IDENTIFIED AS PRODUCT CODE SXPP1A405. IT WAS REQUESTED THAT HSA ASSESS THE FRACTURE MODE OF THE FAILURE. ACCORDING TO THE INFORMATION, IT WAS REPORTED BREAKAGE NEEDLE. THE PRODUCT RECEIVED FOR ANALYSIS WAS SXPP1A405. VISUAL INSPECTION AND METALLURGICAL ANALYSIS WERE PERFORMED ON THE RETURNED PRODUCT. A FRACTURE WAS OBSERVED AT THE SUTURE ATTACHMENT. THE NEEDLE WAS RECEIVED IN TWO PIECES, ONLY ONE OF THE MATING FRACTURE SURFACES WAS EXAMINED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE (SEM) WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS IN ORDER TO DETERMINE THE FRACTURE MODE. THE EVALUATION OF (B)(4) REVEALED THE FRACTURE WAS COMPOSED OF MICROVOID COALESCENCE, WHICH IS EVIDENCE OF A DUCTILE FRACTURE MODE. A CUP-AND-CONE SHAPED FRACTURE AND MACROSCOPIC PLASTIC DEFORMATION OBSERVED COINCIDENTAL TO THE FRACTURE PROVIDE ADDITIONAL EVIDENCE THAT THE FAILURE WAS INDUCED BY MECHANICAL DEFORMATION LEADING TO DUCTILE OVERLOAD. THIS WAS A DUCTILE FRACTURE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT#: (B)(4). H6 COMPONENT CODE: G07002 - DEVICE NOT RETURNED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803, PART 4 SUBPART B. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION WAS REQUESTED, AND THE FOLLOWING WAS OBTAINED VIA: LOT NUMBER? UNK. PLEASE PERFORM AND DOCUMENT THE FOLLOW UP ATTEMPT FOR PRODUCT RETURN. PLEASE DOCUMENT THE SHIPMENT TRACKING NUMBER IN NOTES OR RMAO SECTIONS. THE DEVICE HAS BEEN RECEIVED AT SUKAGAWA AND WILL BE SHIPPED. PLEASE CHECK RMAO. TRACKING NUMBER: (B)(4). PROVIDE THE SOURCE OR NAME AND TITLE OF EXTERNAL PERSON PROVIDING ANSWERS TO FOLLOW-UP (EXTERNAL PERSON SUBMITTING ANSWERS TO SALES REP OR LOC/BQ). (B)(6). THE MANUFACTURING RECORDS COULD NOT BE REVIEWED AS THE BATCH/LOT NUMBER IS UNKNOWN. D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS BROKEN AT THE PROXIMAL END DURING USE AND USE WAS STOPPED. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO PIECES WERE LEFT INSIDE THE PATIENT. IT WAS NOT A CONTROL RELEASE NEEDLE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2752525 SFXSYM PDS+ UNI VIO 18IN 1 SA CT SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE NEW ETHICON INC.

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown