FDA Adverse Event
Malfunction
Summary report: N
TIMESH STRAIGHT PLATE, 1 X 1, 12MM SHAFT
MDR report key: 2254129
·
Received August 18, 2011
Report
- Report Number
- 2021898-2011-00195
- Event Type
- Malfunction
- Date Received
- August 18, 2011
- Date of Event
- July 14, 2011
- Report Date
- July 19, 2011
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JEY
- PMA / PMN Number
- K973145
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE EYELET OF THE PLATE WAS BROKEN AND BENT OPEN. THE TAB NORMALLY ATTACHED TO THE PLATE WAS NOT RETURNED. IT IS UNKNOWN WHEN OR HOW THIS DAMAGE OCCURRED. PER THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE, BREAKAGE OF TIMESH COMPONENTS IS A KNOWN COMPLICATION. A REVIEW OF THE MANUFACTURING RECORDS SHOWED NO ANOMALIES. NO IMPACT TO THE PATIENT WAS REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED TO MEDTRONIC NEUROSURGERY THAT THE DOCTOR FOUND FISSURES ON THE EDGE OF THE CONNECTOR DURING SURGERY.