FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 2254102 · Received September 19, 2011

Report

Report Number
2939301-2011-09149
Event Type
Malfunction
Date Received
September 19, 2011
Report Date
September 12, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.THE 510(K)# IS K082590.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON OCTOBER 24, 2011 WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A BATTERY CONTACT CORROSION AND A DEFECTIVE PCB BOARD. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE BATTERY CONTACT ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC.

Patients

Seq Age Sex Outcome Treatment
1