ENDOWRIST
Report
- Report Number
- 2955842-2025-30462
- Event Type
- Malfunction
- Date Received
- July 18, 2025
- Date of Event
- June 6, 2025
- Report Date
- June 23, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874119808
- PMA / PMN Number
- K214095
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- 003
Narratives
A RETURN MATERIAL AUTHORIZATION (RMA) WAS ISSUED TO EVALUATE THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT, HOWEVER, INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE INSTRUMENT FOR EVALUATION. THEREFORE, THE ROOT CAUSE OF THE REPORTED FAILURE MODE HAS NOT YET BEEN DETERMINED.
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO REPRODUCE/CONFIRM THE REPORTED COMPLAINT. THE INSTRUMENT WAS FOUND TO HAVE A BROKEN CONDUCTOR WIRE WITHIN THE BIPOLAR YAW PULLEY, BETWEEN THE WIRE CRIMP ON THE GRIP AND THE SILICONE POTTING, WHICH SEALS THE WIRE ENTRANCE TO THE YAW PULLEY. THE INSTRUMENT FAILED THE ELECTRICAL CONTINUITY TEST, CONFIRMING A COMPLETE BREAK IN THE WIRE. NO THERMAL DAMAGE WAS FOUND ON THE INSTRUMENT. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO COMPONENT SUSCEPTIBILITY TO DAMAGE THROUGH EXTERNAL COLLISIONS, DURING USE, OR DURING REPROCESSING. THIS ISSUE CAN BE RESOLVED BY USING AN ALTERNATE INSTRUMENT TO COMPLETE THE PROCEDURE.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY RADICAL EXTRAPERITONEAL WITH LYMPHADENECTOMY SURGICAL PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS OBSERVED TO HAVE A DAMAGED CONDUCTOR CABLE. THERE WAS NO TISSUE DAMAGE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE DISTRIBUTOR AND OBTAINED THE FOLLOWING INFORMATION: THE BLACK CABLE WAS COMPLETELY BROKEN. THERE WAS LOSS OF CAUTERY DUE TO THE FAILURE. THERE WERE NO SIGNS OF ARCING OR THERMAL DAMAGE. THE INSTRUMENT DID NOT COLLIDE WITH ANOTHER INSTRUMENT DURING THE PROCEDURE AND THERE WAS NO SIGN OF ARCING. THE DISTRIBUTOR DID NOT KNOW WHAT OTHER INSTRUMENTS WERE IN USE WHEN THE ISSUE OCCURRED.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 878505 | ENDOWRIST | FENESTRATED BIPOLAR FORCEPS | NAY | INTUITIVE SURGICAL, INC | 471205-19 | K13241121 0012 | 00886874119808 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | DA VINCI INSTRUMENTS AND ACCESSORIES |