FDA Adverse Event Malfunction Summary report: N

TRUSTEEL

MDR report key: 22539168 · Received July 18, 2025

Report

Report Number
3003442380-2025-11681
Event Type
Malfunction
Date Received
July 18, 2025
Date of Event
June 21, 2025
Report Date
October 6, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018488
PMA / PMN Number
K041545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (3003442380-2025-11681), WAS SUBMITTED ON 18-JUL-2025. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 26-NOV-2024. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010326, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 06-OCT-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010326". THE COUNT OF COMPLAINT IS 0 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010326 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 98 AND MANUFACTURED IN THE MACHINE MULTIVAC14 ON 26/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. GLUE-CONNECTOR LOT: THE LOT 4K05519 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE MP03 ON 14/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4K05776 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE MP03 ON 14/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4K05777 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE MP03 ON 24/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4L03430 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE MP03 ON 25/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT 4L03431 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE MP03 ON 26/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025 THE BLOCKAGE WAS AT THE SITE. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE, ADDITIONAL PATIENT OR EVENT DETAILS WERE RECEIVED WHICH HAVE BEEN ADDED IN H11.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
821544 TRUSTEEL UNO CONTACT DETACH G29 80/6TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002835 6010326 05705244018488

Patients

Seq Age Sex Outcome Treatment
1 66 YR Female