FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2253842
·
Received August 11, 2011
Report
- Report Number
- 1824206-2011-04241
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- July 29, 2011
- Report Date
- July 29, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND THE BRAKES WERE NOT HOLDING AND WOULD RATCHED FROM SIDE TO SIDE. BED FOUND OUTSIDE ICU IN HALLWAY. REPLACED ALL BRAKE AND STEER CASTERS TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
BRAKES WERE NOT HOLDING AND WOULD RATCHET FROM SIDE TO SIDE. NO INJURY ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | P1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |