FDA Adverse Event Malfunction Summary report: N

AFFINITY 3

MDR report key: 2253837 · Received August 11, 2011

Report

Report Number
1824206-2011-04263
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE 37 PIN CONNECTOR WAS BROKEN ON THE UTV RELAY J-BOX CIRCUIT BOARD. THE TECHNICIAN REPLACED THE J-BOX CIRCUIT BOARD TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BED IS NOT SENDING A NURSE CALL TO THE NURSES' STATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 3 BIRTHING BED HDD HILL-ROM INC. 3700

Patients

Seq Age Sex Outcome Treatment
1