FDA Adverse Event Malfunction Summary report: N

ADVANCE 2000

MDR report key: 2253805 · Received August 10, 2011

Report

Report Number
1824206-2011-04229
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
July 28, 2011
Report Date
July 29, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THE BRAKES WOULD NOT ADJUST OR HOLD CORRECTLY. REPLACED DAMAGED PLATE AND BEARINGS TO RESOLVE THIS ISSUE. BED LOCATE IN BED SHOP.

Description of Event or Problem · 1

BRAKE MECHANISM ASSEMBLY NOT WORKING CORRECTLY. NO ALLEGED INJURIES BY ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANCE 2000 AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1105

Patients

Seq Age Sex Outcome Treatment
1