FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2253778 · Received August 10, 2011

Report

Report Number
1824206-2011-04198
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
July 28, 2011
Report Date
July 28, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ISOLATED THE PROBLEM TO A DEFECTIVE CPR VALVE. REPLACED THE VALVE TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

FACILITY ALLEGED THAT THE HEAD SECTION WAS DRIFTING DOWN VERY SLOWLY. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900

Patients

Seq Age Sex Outcome Treatment
1