FDA Adverse Event Malfunction Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25LPA

MDR report key: 2253698 · Received September 9, 2011

Report

Report Number
3003637274-2011-00050
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 16, 2011
Report Date
August 22, 2011
Manufacturer
KAVO DENTAL GMBH
Product Code
EFA
PMA / PMN Number
K073478
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE PRODUCT DID SHOW THAT THE HANDPIECE HAD A MEDIUM DEBRIS LEVEL, THE BEARINGS HAVE BEEN GRINDING, WOBBLING AND FALLING APART. THIS CAUSED THE HEAD TO HEAT UP. THE USER INSTRUCTION STATES THAT A HANDPIECE IN SUCH CONDITION MUSTN'T BE USED. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE. (B)(4). KAVO DENTAL (B)(4) (THE MFR) IS SUBMITTING THE REPORT ON BEHALF OF (B)(4).

Description of Event or Problem · 1

DURING A STANDARD TREATMENT WITH A DENTAL ELECTRICAL HANDPIECE A PT RECEIVED A SMALL BURN ON CHEEK. OFFICE APPLIED A UNK TYPE OF OINTMENT BUT NOTHING WAS PRESCRIBED. NO FURTHER MEDICAL CARE WAS NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25LPA DENTAL HANDPIECE EFA KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Other