FDA Adverse Event Malfunction Summary report: N

MCA LARGE APPLIER

MDR report key: 2253512 · Received September 19, 2011

Report

Report Number
3005075853-2011-03866
Event Type
Malfunction
Date Received
September 19, 2011
Date of Event
August 16, 2011
Report Date
August 30, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDO
PMA / PMN Number
K820837
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS USED DURING AN OPEN COLON RESECTION. THE CLIPS WERE FALLING OUT OF THE JAWS. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MCA LARGE APPLIER CLIP, IMPLANTABLE GDO ETHICON ENDO-SURGERY, LLC. NA H43J14

Patients

Seq Age Sex Outcome Treatment
1