FDA Adverse Event Other Summary report: N

S3 MAST ROLLER PUMP

MDR report key: 2253447 · Received September 9, 2011

Report

Report Number
1718850-2011-00101
Event Type
Other
Date Received
September 9, 2011
Date of Event
August 10, 2011
Report Date
August 11, 2011
Manufacturer
SORIN GROUP DEUTSCHLAND
Product Code
DTQ
PMA / PMN Number
K950990
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MFR DATE WILL BE PROVIDED IN A F/U WHEN AVAILABLE. SORIN GROUP (B)(4) MFRS THE S3 MAST ROLLER PUMP. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP (B)(4) RECEIVED A COMPLAINT REPORTING THAT, DURING THE PROCEDURE, THE VENT PUMP STOPPED AND A 040000 ERROR CODE WAS DISPLAYED. THE ENGINEER POWER CYCLED THE PUMP BUT IT DID NOT REGAIN FUNCTION. THE CLINICIAN ELECTED TO COMPLETE THE OPERATION WITHOUT THE VENT PUMP. AFTERWARDS, THE ENGINEER STARTED THE PUMP, RAN IT FOR A PERIOD OF TIME AND THEN ANOTHER ERROR CODE WAS DISPLAYED, 08000, AND THE PUMP AGAIN STOPPED. THERE WAS NO REPORT OF INJURY TO THE PT OR ADDITIONAL MEDICAL INTERVENTION REQUIRED AS A RESULT OF THIS INCIDENT. SORIN GROUP (B)(4) HAS REQUESTED THAT THE PUMP BE RETURNED FOR EVAL. TO DATE NO PRODUCT HAS BEEN RECEIVED. THE INVESTIGATION IS ONGOING. A F/U REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

SORIN GROUP (B)(4) RECEIVED A COMPLAINT REPORTING THAT, DURING THE PROCEDURE, THE VENT PUMP STOPPED AND A 040000 ERROR CODE WAS DISPLAYED. THE ENGINEER POWER CYCLED THE PUMP BUT IT DID NOT REGAIN FUNCTION. THE CLINICIAN ELECTED TO COMPLETE THE OPERATION WITHOUT THE VENT PUMP. AFTERWARDS, THE ENGINEER STARTED THE PUMP, RAN IT FOR A PERIOD OF TIME AND THEN ANOTHER ERROR CODE WAS DISPLAYED, 08000, AND THE PUMP AGAIN STOPPED. THERE WAS NO REPORT OF INJURY TO THE PT OR ADDITIONAL MEDICAL INTERVENTION REQUIRED AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S3 MAST ROLLER PUMP CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS DTQ SORIN GROUP DEUTSCHLAND 50-60-60 NA

Patients

Seq Age Sex Outcome Treatment
1