SUPRAMID BLACK 4/0 (1.5) 45CM DS16
Report
- Report Number
- 3003639970-2025-00358
- Event Type
- Malfunction
- Date Received
- July 17, 2025
- Report Date
- July 23, 2025
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- GAR
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SUMMARY OF INVESTIGATION: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED 11 CLOSED SAMPLES FOR ANALYSIS. WE HAVE CHECKED THE SAMPLES RECEIVED AND THREAD SURFACE IS THE CURRENT AND USUAL ONE, THE COVER IS NOT SEPARATED/DAMAGED. WE HAVE ALSO TESTED THE KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 1.15 KGF IN AVERAGE AND 0.98 KGF IN MINIMUM (EP REQUIREMENTS: 0.51 KGF IN AVERAGE AND 0.15 KGF IN MINIMUM). BATCH MANUFACTURING RECORD: REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. CONCLUSION ROOT CAUSE ANALYSIS: IT HAS NOT BEEN POSSIBLE TO DETERMINE THE ROOT CAUSE BECAUSE THE CLOSED SAMPLES RECEIVED COMPLY USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE CLOSED SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. THE CASE IS CONSIDERED NOT CONFIRMED BY EVIDENCE OF THE CLOSED SAMPLES RECEIVED. CORRECTIVE MEASURES: ACTIONS ON DISTRIBUTED PRODUCT OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.
G4: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. RELATED 510K NUMBER K151165. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED AN ISSUE WITH SUPRAMID SUTURE. THE CLIENT REPORTED THAT THE THREAD BREAKS, THE COATING OF THE THREAD IS COMING OFF. NO FURTHER INFORMATION HAS BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159511 | SUPRAMID BLACK 4/0 (1.5) 45CM DS16 | OTHER SUTURES | GAR | B. BRAUN SURGICAL, S.A. | G0712132 | 625026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |