INFUSOR
Report
- Report Number
- 6000001-2011-25279
- Event Type
- Malfunction
- Date Received
- September 16, 2011
- Date of Event
- August 22, 2011
- Report Date
- August 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING NO SOLUTION IN THE RESERVOIR BECAUSE THE TUBING HAD BEEN CUT TO DRAIN THE FLUID. THE REPORTED CONDITION OF NO FLOW/NON-DELIVERY WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE DEVICE SHOWED NO SIGNS OF BLOCKAGE THAT COULD HAVE CAUSED THE REPORTED CONDITION. THE TUBING WAS RECONNECTED, AND THE RESERVOIR WAS REFILLED WITH DEXTROSE SOLUTION FOR A FUNCTIONAL TEST. AFTER FILL, FLOW WAS READILY OBSERVED AT THE LUER. FLOW CONTINUED WITHOUT STOPPING UNTIL THE SOLUTION WAS EMPTIED FROM THE RESERVOIR. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR HAD DELIVERY STOP DURING PATIENT USE. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THIS CONDITION INTERRUPTED DELIVERY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 11E041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |