COLLEAGUE
Report
- Report Number
- 6000001-2011-25235
- Event Type
- Malfunction
- Date Received
- September 16, 2011
- Date of Event
- July 22, 2011
- Report Date
- September 7, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH AN 814:04 FAILURE CODE WAS FOUND IN THE PUMP'S EVENT HISTORY BUT NOT DUPLICATED DURING PRODUCT EVALUATION. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THIS CONDITION AND NO REPAIR WAS NECESSARY. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH AN 814:04 FAILURE CODE. THE EVENT WAS REPORTED TO HAVE OCCURRED UPON POWER UP IN GENERAL PATIENT WARD. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |