FDA Adverse Event Malfunction Summary report: N

Q-SYTE

MDR report key: 22520463 · Received July 16, 2025

Report

Report Number
1710034-2025-01192
Event Type
Malfunction
Date Received
July 16, 2025
Date of Event
June 22, 2025
Report Date
July 18, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FPA
UDI-DI
00382903851003
PMA / PMN Number
K013621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385100 AND LOT NUMBER 3213081. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD Q-SYTE LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE PATIENT IS A CHRONIC LIVER FAILURE PATIENT WHO REQUIRES LONG-TERM INTRAVENOUS THERAPY VIA THE SUBCLAVIAN VEIN. TO ENSURE THE SAFETY OF THE INFUSION AND PREVENT BLOOD REFLUX, A SEPTUM-TYPE NEEDLELESS INJECTION CAP WAS CONNECTED TO THE RIGHT SUBCLAVIAN VEIN FOR INFUSION THERAPY AT 9:00 A.M. ON (B)(6) 2025, IN THE ICU. IT WAS FOUND THAT THE SEPTUM-TYPE NEEDLELESS INJECTION CAP COULD NOT PREVENT BLOOD REFLUX, SO IT WAS IMMEDIATELY REPLACED WITH A SEPTUM-TYPE NEEDLELESS INJECTION CAP, WHICH FUNCTIONED NORMALLY.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.