ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2251991 · Received September 16, 2011

Report

Report Number
2531779-2011-06859
Event Type
Malfunction
Date Received
September 16, 2011
Report Date
August 17, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 11/17/2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6), 2011 WITH THE FOLLOWING FINDINGS: THE REPORTED COMPLAINT WAS THE PUMP NOT SAVING CARBOHYDRATE INFORMATION PUT INTO THE PUMP DURING BOLUS PROGRAMMING. THE DOWNLOAD HISTORY SHOWED ONLY EZBG AND NORMAL BOLUSES. CARBOHYDRATE ENTRY IS NOT AN OPTION DURING AN EZBG BOLUS THEREFORE THE INFORMATION WOULD NOT SHOW IN THE PUMP HISTORY. NO HISTORY RECORDING ISSUES WERE FOUND DURING TESTING.

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

THE PATIENT'S MOTHER CONTACTED ANIMAS ALLEGING A MEMORY ISSUE WITH THE PUMP. THE REPORTER CLAIMED THE PUMP WAS NOT SAVING CARBOHYDRATE INFORMATION ENTERED INTO THE PUMP WHEN PROGRAMMING BOLUSES. THERE WAS NO REPORTED PATIENT IMPACT AS RESULT OF THE ALLEGED ISSUE. AT THE TIME OF TROUBLESHOOTING, CUSTOMER SUPPORT NOTED THAT BOLUSES ON DOWNLOAD HISTORY FROM (B)(6) 2011 SHOWED THAT EZBG MENU FEATURE WAS USED ON (B)(6) 2011, (B)(6), 2011 AND (B)(6) 2011. DURING REVIEW OF PUMP HISTORY, CUSTOMER SUPPORT NOTED THAT THE INFORMATION LISTED IN THE DOWNLOAD HISTORY CORRELATED WITH THE INFORMATION NOTED IN THE PUMP HISTORY. HOWEVER, THE REPORTER CLAIMED THE PATIENT USES EZCARB MENU 95% OF THE TIME AND INFORMATION WAS MISSING FROM THE DOWNLOAD HISTORY REPORT AND PUMP'S HISTORY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED COMPLAINT WAS UNRESOLVED AT THE TIME OF TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 18 YR