FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2251972 · Received September 16, 2011

Report

Report Number
1423500-2011-12344
Event Type
Malfunction
Date Received
September 16, 2011
Date of Event
August 25, 2011
Report Date
August 25, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) OCCURRED DURING INITIAL DRAIN WAS CONFIRMED. THE CAUSE WAS USE ERROR WAS AN OPEN CLAMP. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4).A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 1 OF 5. THE HOME PATIENT (HP) STATED AN UNUSED LINE CLAMP WAS OPEN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED ANY UNUSED LINES MUST BE CLAMPED CLOSED. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 74 YR HOMECHOICE